Teledyne-API Model 101E/102E
H2S & TRS Analyzers


The Models 101E/102E provide a dependable, accurate and convenient method of measuring hydrogen sulfide or Total Reduced Sulfur (TRS) in levels commonly required for Ambient Air monitoring, offering ranges from 0-50 ppb to 0-10 ppm. Both analyzers convert sulfur gases to sulfur dioxide and measure concentrations using fluorescence technology.
Model 100E
The M101E is equipped with an internally mounted low temperature converter which converts H2S at a closely controlled temperature setting of 315 °C, leaving other gases unaffected.
Model 102E
The M102E uses a high temperature converter, mounted externally to the analyzer, and operates at a temperature of 850°C. TRS components (typically, H2S, methyl mercaptan, dimethyldisulfide and methyl-disulfide) are all converted to SO2 at this temperature, with efficiency greater than 98%.
Sample gas drawn into the analyzer first passes through a scrubber which removes all traces of sulfur dioxide. The converter then converts the components of interest to SO2. The resulting concentration for SO2 is then measured by fluorescence. In the basic configuration, the analyzer may be programmed to read H2S concentration (101E) or TRS (102E) only. However, a switching option, showing H2S/TRS and SO2 readings alternately, is available. Cycle times may be adjusted, and each reading is truly independent, having its own calibration slope and offset. A sample and hold circuit provides continual analog outputs for each channel. Calibration is simplified by the optional zero and span
valves, which allow direct entry of calibration gases, or the optional permeation tube may be used to provide span checks. The analyzers feature auto zero which provides excellent zero stability.
Specifications
Ranges: 0-50 ppb to 0-20,000 ppb full scale, user selectable. Dual ranges and auto ranging supported
101E Ranges: 0 – 50 ppb to 0 – 10 ppm H2S or up to 20 ppm SO2;
102E Ranges: 0 – 50 ppb to 0 – 10 ppm TRS or 20 ppm SO2
(Note: For ranges higher than 0-10 ppm, please consult Teledyne API)
Range Selection: Ranges are user selectable, and independent ranges are available for both components. Auto and dual ranging are available.
Zero Noise: 0.2 ppb (RMS)
Span Noise: < 0.5% of reading (RMS) above 50 ppb
Lower Detectable Limit (LDL): < 0.4 ppb
Zero Drift: < 0.5 ppb / 24 hrs, 1 ppb / 7 days
Span Drift: < 0.5 % full scale / 24 hrs, 1% full scale / 7 days
Lag Time: 20 seconds
Rise and Fall Time: < 120 seconds
Linearity: 1% of full scale
Sample Flow Rate: 650 cc/min ±10 %
Converter Temperature: 101E: 315°C; 102E: 850°C
Converter Location: 101E: Inside analyzer; 102E: Remote, may be rack mounted
Dimensions (HxWxD): 7” (178 mm) x 17” (432 mm) x 23.5” (597 mm) for analyzer.
Model 102E requires same size for converter
Weight: 101E: 41 lbs (18.3 kg); 102E: 53 lbs (24 kg)
Power: 100v, 50/60 Hz (3.25A); 115V, 60 Hz (3.0A): 220-240V, 50/60 Hz (2.5A)
Pump: Internal
Analog Outputs: 10V, 5V, 1V, 0.1V, selectable
Recorder Offset: ±10%
RS232: Standard DB9 connector
Status (Digital) 8 outputs, 6 inputs (opto-isolated), 4 alarm outputs (optional)
Current Output: Optional 4-20 mA isolated outputs (select up to 3 channels)