Rental/Sales
800 399-4CEM
(800 399-4236)

|
 Teledyne-API Model 300EM
Gas Correlation
CO Analyzer


The Model 300EM measures carbon monoxide by comparing infrared energy absorbed by a sample to that absorbed by a reference gas according to the Beer-Lambert law. This is accomplished with a Gas Filter wheel which alternately allows a high energy light source to pass through a CO filled chamber and a chamber with no CO present. The light then travels through the folded path sample cell. This technique allows near interferent free operation.
The M300EM design produces superior zero and span stability and a high signal-to-noise ratio, allowing excellent sensitivity.
Multi-tasking software gives real time indication of numerous operating parameters and provides automatic alarms if diagnostic limits are exceeded. Built-in data acquisition and internal memory allows logging multiple parameters including average and instantaneous values, calibration data, and operating parameters. Stored data can be retrieved via the RS-232 port, optional Ethernet port, or from the front panel.
The Model 300EM features rugged construction and is designed to perform with a minimum of attention. Regular maintenance is greatly simplified through the use of modularized construction.
Also Available: Teledyne-API Model 300EM w/ CO2
Specifications
| Ranges: |
0-5 ppm to 0-5,000 ppm full scale, user selectable; Dual ranges and auto ranging supported |
| Units: |
Dual ranges and auto ranging supported
ppm, mg/m3 |
| Zero noise: |
< 0.1 ppm |
| Span noise: |
< 0.1 ppm (RMS) |
| Lower Detectable Limit (LDL): |
0.2 ppm |
| Zero drift: |
< 0.5 ppm/24 hours, < 1 ppm/7 days |
| Span drift: |
< 0.5% of reading/24 hours, 1% of reading/7 days |
| Lag time: |
10 seconds |
| Rise and Fall Time: |
< 60 seconds to 95% |
| Linearity: |
1% of full scale |
| Precision: |
1.0% of reading |
| Sample Flow Rate: |
800 cm3/min ±10% |
| Operating Temperature Range: |
5 - 40° C (with EPA equivalency) |
| Dimensions (HxWxD): |
7” (178 mm) x 17” (432 mm) x 23.5” (597 mm) |
| Weight: |
40 lbs (24 kg) |
| Power: |
100V - 120V, 220V - 240V, 50/60 Hz, 250W |
| Analog outputs: |
10V, 5V, 1V, 0.1V, selectable |
| Recorder Offset: |
±10% |
| Serial Outputs: |
Serial Port 1: RS-232 (DB-9M)
Serial Port 2: standard RS-232 or optional RS-485 (DB-9F), Ethernet |
| Status (Digital): |
8 outputs, 6 inputs (opto-isolated), 4 alarm outputs (optional) |
| Current Output: |
Optional 4-20mA isolated outputs, select up to three channels |
NOTE: The values expressed above are in accordance with EPA definitions.
All error specifications are based on constant conditions.
For availability and recommendations, please call us at 800.399.4236 or email a Sales Engineer.
Or click on the RFQ link to submit project details.
|