Teledyne-API Model 803E
CO2/O2 Paramagnetic Analyzer


The NEW 803E uses the proven paramagnetic O2 measurement principle, coupled with a state of the art CO2 NDIR sensor. These are combined with the TAPI E series advanced microprocessor based electronics to provide a high performance analyzer giving accurate,
dependable measurements of percent level Oxygen and Carbon Dioxide. This combination product is ideally suited for measuring extractive source emissions, and both paramagnetic O2 and NDIR CO2 are reliable tools used extensively for RATA’s. The 803E provides a cost effective solution for both markets.
All TAPI E-Series include an extensive built-in data acquisition capability using the analyzer's internal memory. This allows the logging of multiple parameters including averaged and instantaneous concentration values, calibration data and operating parameters such as flow and pressure.
Stored data are easily retrieved through the serial port or optional no-cost Ethernet port and viewed via our APIcom software or from the front panel, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.
The Model 803E combines light-weight, rugged construction, ease of use, powerful diagnostics, modular design and outstanding performance to yield the ideal tool for today’s air monitoring requirements.
Specifications
CO2:
Ranges: 0-20% CO2
Zero Stability: < 0.1% CO2 in 24 hours
Span Noise: < 0.1% RMS CO2
Span Stability: < 0.2% CO2 in 24 hours
Accuracy: 1.5% CO2 of range + 2% of reading
Linearity: < 0.5% full scale CO2
O2:
Ranges: 0-1% to 0-100% user selectable O2
Zero Stability
Short Term: < 0.1% O2 in 24 hours
Long Term: < 0.2% O2 in 7 days
Span Noise: < 0.05% O2
Minimum Detectable Change (MDC): < 0.1% O2
Linearity: < 0.1% O2
Repeatability < 0.01% O2
Lag Time: 10 seconds
Rise and Fall Time: < 60 seconds to 95%
Sample Flow Rate: Max/Min 130ml ± 20ml/min
Operating Temperature Range: 5 - 40˚C
Humidity Range: 0-95% RH
Pressure Range: 25-31 in Hg
Dimensions (HxWxD): 7” (178 mm) x 17” (432 mm) x 23.5” (597 mm)
Weight: 28 lbs (12.7 kg)
Power: 100-120V, 220-240V 50/60Hz
Analog Outputs: 10V, 5V, 1V, 100mV, user selectable
Recorder Offset: ±10%
Serial Outputs: Format (RS232/485, Ethernet, Multi-drop)
One RS-232 port (DB-9 male), one RS-232 or RS-485 User Selectable (DB-9 female)
Status (Digital): 8 outputs, 6 inputs (opto-isolated), 6 alarm outputs (optional)
Current Output: Optional 4-20mA, select up to three channels
For availability and recommendations, please call us at 800.399.4236 or email a Sales Engineer.
Or click on the RFQ link to submit project details.