QuickShip Rentals

Parts & Repairs
Product Catalog
Order Tracking
Ask Cherokee!

CEM Systems
CEM Services

CEM Consulting
CEM Resources

About Us
Contact Us
Home

Contact Us:
800 399-4CEM
(800 399-4236)
Sales email


info request form

RFQ form




Add Cherokee Instruments
to your

Google Bookmarks

Continuous Emissions Monitors

Thermo Environmental Model 48C/H
Carbon Monoxide Analyzer


The Thermo Model 48 CO analyzer uses a gas filter correlation/non-dispersive infrared detector. The gas filter correlation technique reduces instrument cross-sensitivity to similar gases. Calibration values are adjusted via the zero and span thumb wheels or microprocessor (depending on the model). The Model 48C offers ranging up to 30,000 ppm and the Model 48H offers ranging up to 50,000 ppm.

Manufacturer Thermo Environmental Instruments
Analysis Method Gas Filter Correlation
Range 0 - 30,000 ppm and 0-50,000 ppm
Precision 1% of reading or 0.05 ppm
Zero Drift < 0.1 ppm in 24 hrs.
Span Drift + 1% full scale in 24 hrs.
Sample Flow Rate 1 liter / min.
Analog Output Selectable voltage and RS-232 standard
Power 115VAC
Dimensions 16.75” x 8.62” x 23”
Weight 45 lbs

API Model 300 Carbon Monoxide Analyzer




The API Model 300 CO analyzer uses a gas filter correlation/non-dispersive infrared detector. The gas filter correlation technique reduces instrument cross-sensitivity to similar gases. Calibration values are adjusted via the microprocessor controls. The Model 300 offers excellent sensitivity and has user defined ranging of up to 1,000 ppm.

Manufacturer Advanced Pollution Instrumentation, (API)
Analysis Method Gas Filter Correlation / NDIR
Range 0 - 1000 ppm Zero Noise 0.025 ppm
Response Time < 60 seconds
Linearity 1% of full scale
Zero Drift < .1 ppm (24hrs)
Span Drift < 1% full scale
Sample Flow Rate
1 liter / min.
Analog Output 0 to 1, 5, 10 VDC and 4 to 20 mA
Power 115 VAC Dimensions 17” x 7” x 25”
Weight 50 lbs

California Analytical Model 300P O2/CO2/CO Analyzer
The CAI Model 300P is a multiple detector O2/CO2/CO analyzer that employs a paramagnetic sensor for O2 analysis and an NDIR detector for CO2 and CO analysis. The NDIR detectors employ optical filtration to negate cross-interferences. The instrument is provided with an internal adjustable sample pump, front panel zero and span potentiometers, and a wide variety of range selections.

Ranges

O2: 0-1, 5, 10, 15, 25, 40 and 100% by volume (specify)
CO2: 0-20, 40% by volume (typical)
CO: 0-200, 500, 1000, 2500 ppm (depends on unit)

Detection Method

Paramagnetic (O2), NDIR (CO2, CO)

Sample Gas Input

Single 1/4-inch Tube; 0.5 to 2 lpm

Analog Outputs

Independent 0-10 Vdc, 4-20mA (others upon request)

Power Requirement

110 VAC/60 Hz

Shipping & Weight

32x25x14 – 70 lbs

Thermo Environmental Model 48C CO Analyzer
The Thermo Model 48C CO analyzer utilizes the technique of gas filter correlation coupled with an NDIR detector. This technique minimizes interferences from other gases. The instrument is microprocessor controlled and is provided with an internal sample pump, front panel zero and span “push button” adjustments, and a wide variety of range selections. USEPA designated (RFCA-0981-054)

Ranges

0-1 to 0-10000 ppm (pre-defined and/or user defined)

Detection Method

GFC/NDIR

Sample Gas Input

1/4-inch Tube; 1 lpm

Analog Outputs

0-1, 5, 10 Vdc, 4-20mA (please specify)

Power Requirement

110 VAC/60 Hz

Shipping & Weight

32x25x14 – 71 lbs

Thermo Environmental Model 48C-HL CO Analyzer
The Thermo Model 48C-HL CO analyzer is a high concentration level CO analyzer that utilizes the technique of gas filter correlation coupled with an NDIR detector. This technique minimizes interferences from other gases. The instrument is microprocessor controlled and is provided with an internal sample pump, front panel zero and span “push button” adjustments, and a wide variety of range selections.

Ranges

0-50 to 0-20000 ppm (pre-defined and/or user defined)

Detection Method

GFC/NDIR

Sample Gas Input

1/4-inch Tube; 1 lpm

Analog Outputs

0-1, 5, 10 Vdc, 4-20mA (please specify)

Power Requirement

110 VAC/60 Hz

Shipping & Weight

32x25x14 – 71 lbs

API Model 300 Carbon Monoxide Analyzer
The API Model 300 CO analyzer uses the technique of gas filter correlation coupled with an NDIR detector. The gas filter correlation technique minimizes instrument cross-sensitivity from other gases. The instrument is microprocessor controlled and is provided with an internal sample pump, front panel zero and span “push button” adjustments, and a wide variety of range selections. USEPA designated (RFCA-1093-093)

Ranges

0-1 to 0-1000 ppm (user defined)

Detection Method

GFC/NDIR

Sample Gas Input

1/4-inch Tube; 1 lpm

Analog Outputs

0-1, 5, 10 Vdc, 4-20mA (please specify)

Power Requirement

110 VAC/60 Hz

Shipping & Weight

32x25x14 – 75 lbs

API Model 300EM Carbon Monoxide Analyzer
The API Model 300EM CO analyzer uses the technique of gas filter correlation coupled with an NDIR detector. The gas filter correlation technique minimizes instrument cross-sensitivity from other gases. The instrument is microprocessor controlled and is provided with an internal sample pump, front panel zero and span “push button” adjustments, and a wide variety of range selections.

Ranges

0-3 to 0-3000 ppm (user defined)

Detection Method

GFC/NDIR

Sample Gas Input

1/4-inch Tube; 1 lpm

Analog Outputs

0-1, 5, 10 Vdc; 4-20mA (please specify if desired)

Power Requirement

110 VAC/60 Hz

Shipping & Weight

32x25x14 – 75 lbs

API Model 200EH NO/NO2/NOx Analyzer
The API Model 200EH is a microprocessor-controlled chemiluminescence NOx analyzer that can simultaneously measure NO, NO2, and total NOx concentrations. A high temperature, stainless steel converter is included as standard, but the unit can be configured with a low temperature, molybdenum converter for ammonia injection applications. An external vacuum pump (included) provides vacuum for the reaction chamber and dry, filtered air for ozone generation.

Ranges

0-5 to 0-5000 ppm (user defined)

Detection Method

Chemiluminescence                                                

Sample Gas Input

1/4-inch Tube; 0.75 to 1 lpm

Analog Outputs

Independent 0-1, 5, 10 Vdc; 4-20mA
(please specify if desired)

Power Requirement

110 VAC/60 Hz

Shipping & Weight
(2 pieces)

Vacuum Pump = 14x15x16 – 25 lbs
Analyzer = 32x25x14 – 72 lbs

:: Back to Continuous Emissions Monitors ::

Our Affiliates

Teledyne API

Air Products & Chemical Gases
Apex Instruments, Inc.
AMATEK Western Research (Bovar)
Baldwin Environmental
California Analytical Instruments, Inc.
Cherokee Instuments
Emission Measurement Technology
Graseby / Andersen / General Metal Works
MSA Baseline
O’Brien Corporation
Pacwill Environmental
Spectra Gases, Inc.
STRATA
Telegan Gas Monitoring, LTD
Thermo Electron
Universal Analyzers


cems air quality emissions testing
CEM Systems & Services
 

Contact Us | Privacy Statement
©2008 Cherokee Instruments, Inc.