VIG Model-210 Heated Methane/Ethane/NonMethane-Ethane
Total Hydrocarbon Analyzer (NMHC)










The VIG Industries, Inc. Model-210 is a microprocessor based, oven heated methane /ethane/nonmethane-ethane/total hydrocarbon gas analyzer designed for high accuracy, sensitivity and stability. The Model-210 uses two independent flame ionization detectors (FIDs), one to measure total hydrocarbons and the second coupled with a GC Column for the separation of the methane, ethane and nonmethane-ethane components. A sample is fed to the analyzer via an internal heated pump to the first FID for a real time total hydrocarbon reading. A portion of the sample is trapped and pushed through a column to separate the methane and ethane components and then to the second FID. Any remaining sample in the column is back flushed through the column to obtain the nonmethane-ethane components. All components that come in contact with the sample through analysis are maintained in a temperature-controlled oven to prevent condensation, and to provide repeatable, reliable performance in the analysis of a wide variety of hydrocarbon concentrations in gaseous mixtures or in ambient air.


The VIG Model 210 THC/Methane/Ethane analyzer employs two heated flame ionization detectors. One detector measures total hydrocarbons continuously; the second detector is coupled to a GC column for the separation of the methane, ethane and non-methane-ethane components. A sample is injected into the column and the methane component passes through to the detector; the ethane component then follows through to the detector for analysis. The remaining sample is then back flushed through the column and to the detector to obtain the non-methane/ethane portion of the sample. Sample loop duration is approximately 3 minutes.
Ranges

0-10, 100, 1000, 10000 ppm OR
0-100, 1000, 10000, 100000 ppm (specify)

Detection Method

FID and FID/GC                                        

Sample Gas Input

1/4-inch Tube; 4 lpm

Analog Outputs

0-10 Vdc, 4-20mA (please specify if desired)

Options
• Zero and calibration solenoids with software
• RS-232 interface
• Internal combustion air supply



Specifications

Analysis Method

Flame Ionization Detector (FID)

Separation Method

Column

Precision

+/- 1% of Full Scale

Response Time

  • Total Hydrocarbons - Within 5 seconds to 90% of final reading
    (continuous real time reading)
  • Methane - Approximately 40 seconds,
    updated every 3 minutes
  • Ethane - Approximately 55 seconds,
    updated every 3 minutes
  • Non methane - Approximately 90 seconds,
    updated every 3 minutes

Zero Drift

+/- 1% of Full Scale Per 24 Hours

Span Drift

+/- 1% of Full Scale Per 24 Hours

Linearity

Within 1%

Oxygen Synergism

Within 1% of Selected Range

Ranges

Standard Ranges (4 Ranges per amplifier, two amplifiers per analyzer,
one amplifier for total hydrocarbons,
one amplifier for methane, ethane and non methane-ethane)
  • 0-100, 1000, 10000, 100000ppm (LDL 0.1ppm)
  • 0-10, 100, 1000, 10000ppm (LDL 0.01ppm)
 

Display

Backlit Digital / Graphical LCD

Outputs

0-1, 5, or 10 Volts and 4-20mA.

Alarms

Flame Out & Concentration, Latching

Safety

Flame Out Indicator with Fuel Shut-Off System

Oven Temperature

Operating Temperature 275°F, Adjustable 200 to 300°F

Sample Pump

Heated 4 Liters/Minute standard or Heated 14 Liters/Minute optional
Ignition Automatic (can be set to manual by operator)
Glow Plug Main and spare glow plug, selectable by switch on back panel
Dimensions 19" front panel, 16.75" wide chassis, 18" deep chassis,
21" deep with fittings and handles, 9" high
Weight 55 to 65 pounds depending on options