The API Model T200H is a microprocessor-controlled chemiluminescence analyzer that can simultaneously measure NO, NO2, and total NOx at 0-5 ppm / 0-5,000 ppm concentrations. A high temperature, stainless steel converter is included as standard, but the unit can be configured with a low temperature, molybdenum converter for ammonia injection applications. An external vacuum pump (included) provides vacuum for the reaction chamber and dry, filtered air for ozone generation.
Teledyne-API T200H Data Sheet
H2S Ranges: Min: 0-50 ppb Full scale Max: 0-10 ppm Full scale
SO2 Ranges: Up to 0-20 ppm Full scale (selectable, independent ranges and auto ranging supported) Measurement Units: ppb, ppm, μg/m3, mg/m3 (selectable) Zero Noise: < 0.2 ppb (RMS) Span Noise: < 0.5% of reading (RMS) above 50 ppb Lower Detectable Limit: 0.4 ppb Zero Drift: < 0.5 ppb/24 hours Span Drift: < 0.5% of full scale/24 hours Lag Time: 20 seconds Rise and Fall Time: < 120 seconds to 95% Linearity: 1% of full scale Precision: 0.5% of reading above 50 ppb Sample Flow Rate: 650 cm3/min ±10% Electrical Specifications Power Requirements: 100V-120V, 220V-240V, 50/60 Hz Analog Output Ranges: 10V, 5V, 1V, 0.1V, (selectable) Recorder Offset: ±10% Communication Specifications Included I/O: 1 x Ethernet: 10/100Base-T 2 x RS232 (300-115,200 baud) 2 x USB device ports 8 x opto-isolated digital outputs 6 x opto-isolated digital inputs 4 x analog outputs
|